Functions6.19 [3.05] Maintenance & Diagnostics > Diagnostics valuesOperating with the Local User InterfaceOperating Manual, 06/2017, A5E31930478-05 219Only ULTRAMAT 7: predictive self-diagnosticsTable 6- 58 ULTRAMAT 7: [3.05.2] Predictive self-diagnosticsParameter DescriptionSource intensity High alarm limit Maximum positive deviation of the source intensity (in %). Amaintenance alarm is signaled on a high limit violation.High warning limit Maximum positive deviation of the source intensity (in %). Amaintenance demanded is signaled on a high limit violation.Current deviation Percentage deviation of the current intensity I0 from thenormalized intensity I0 norm .Low warning limit Maximum negative deviation of the source intensity (in %). Amaintenance demanded is signaled on a low limit violation.Low alarm limit Maximum negative deviation of the source intensity (in %). Amaintenance alarm is signaled on a low limit violation.Intensity I 0 TC Source intensity I 0 after temperature compensation6.19.3 [3.05.3] Electric parametersOverviewIn this menu you access information/current values that you can use to evaluate the supplystatus of the device. The menu shows the modules installed in the device for which electricalparameters are available. This includes analyzer modules and any installed option modules.The menu can include up to three navigation lines, depending on the features:You only have read access.Navigation lines in the menu [3.05.03]Navigation number 1) Module name 2) Component/ moduletype 3) Display details ►1) AM1 → "1.", AM2 → "2.", OM2 → "4."; "3" currently not assigned.2) Depending on features: AM1, AM2, OM23) Gas component or module type, for OM2, e.g., "AI/DI"