4 Appendices68 Keysight U2722A/U2723A User’s GuideAppendix B: Time-out SettingsThe following procedure provides the instructions on how to set the time-outvalue in the Keysight IO Libraries.1 To launch your Keysight Connection Expert, go to Start > All Programs >Keysight IO Libraries Suite > Keysight Connection Expert.2 On the Instrument I/O on this PC panel, select your device in the USBexpanded list and right-click.3 Select Send Commands To This Instrument and the Keysight Interactive IOwindow will appear.4 Go to Interactive IO > Options and the Options dialog box will appear.5 Configure the time-out value based on the table provided below.Table 4-2 Time-out settingsPoints Tint (ms) Time-out range (ms)1 to 10 1 to 100 Std IO 5000101 to 32767 Assign 110% of t11 of 50 1 to 50 Std IO 500051 to 32767 Assign 120% of t51 to 100 1 to 10 Std IO 500011 to 32767 Assign 120% of t101 to 5001 to 5 Std IO 50006 to 500 Assign 130% of t501 to 32767 Assign 110% of t501 to 1000 1 to 500 Assign 310% of t501 to 32767 Assign 120% of t1001 to 4096 1 to 500 Assign 310% of t501 to 32767 Assign 110% of tNOTE – Expected time, t = Points× Tint– The Points is set using the SENSe:SWEep:POINts command and the Tint is setusing the SENSe:SWEep:TINTerval command.