297 ADAM-4000 Series User ManualAppendix A Technical SpecificationsTable A.12: ADAM-4018M Range Accuracy for ThermocoupleRange Code (Hex) Input Range Typical Accuracy Maximum Error Units0E J thermocouple0 to760 °C ±1.0 ±1.5 °C0F K thermocouple0 to 1370 °C ±1.0 ±1.5 °C10 T thermocouple-100 to 400 °C ±1.0 ±1.5 °C11 E thermocouple0 to 1000 °C ±1.0 ±1.5 °C12 R thermocouple500 to 1750 °C ±1.2 ±2.5 °C13 S thermocouple500 to 1750 °C ±1.2 ±2.5 °C14 B thermocouple500 to 1800 °C ±2.0 ±3.0 °CNote! Because the CJC sensor of ADAM-4018M is located in the side of chan-nel 0 to 4, the measurement will have the difference ± 1 °C betweenchannel 0 ~ 4 and channel 5 ~ 7.