A p p e n d i x F A c o u s t i c O u t p u t R e f e r e n c eI n s t r u c t i o n s f o r U s e F - 6 7Acoustic Output Reporting Table – Track 3, FDA 510(k) and IEC 60601-2-37(Per transducer/mode that exceeds MI or TI value of 1.0)Transducer model: 9EVF4 Operating mode: M-ModeMI TIS TIB TICIndex Label Scan Non-scan Non-scanAaprt ≤1 Aaprt >1Maximum Value 1.49 - 0.31 - 0.51 0.46Associated Acoustic ParametersFDA IEC UnitsPr.3 pra (MPa) 3.34W o P (mW) - 13.47 13.47 13.47min of [W .3(z1),ITA.3(z1)]Min. of [Pα(zs ),Ita.α(zs)](mW) -z1 zs (cm) -zbp zbp (cm) -zsp zb (cm) 1.40z@PII .3max z @ max. Ipi,α (cm) 1.50d eq(zsp) deq(zb) (cm) 0.37f c fawf (MHz) 5.05 - 4.85 - 4.85 4.85Dim. of A aprt X (cm) - 0.69 - 0.69 0.69Y (cm) - 0.60 - 0.60 0.60Other Information PD td (μsec) 0.49PRF prr (Hz) 600.00Pr @ PII max Pr @ max. Ipi (MPa) 4.16d eq@ PII max deq@ max. Ipi (cm) 11.31Focal Length FLx (cm) - 10.53 - 10.53FLy (cm) - 3.00 - 3.00Ipa.3@ MImax Ipa.3@ MI max (W/cm2) 460.79OperatorControl Frequency (MHz) 4.00 - 4.00 - 4.00 4.00Focus (cm) 2.00 - 10.53 - 10.53 10.53Gate Size (cm) 1.00 - 1.00 - 1.00 1.00Sample Rate (Hz) 600.00 - 600 - 600 600a This Index is not relevant to this operating mode.b This transducer is not intended for transcranial or neonatal cephalic uses.c This formulation for TIS is less than that for an alternate formulation in this mode.# No data is provided for this operation condition since the maximum index value is not reported for the reason listed.Acoustic Output Reporting Table – Track 3, FDA 510(k) and IEC 60601-2-37(Per transducer/mode that exceeds MI or TI value of 1.0)Transducer model: CW2 Operating mode: CW DopplerMI TIS TIB TICIndex Label Scan Non-scan Non-scanAaprt ≤1 Aaprt >1Maximum Value 0.08 - 1.60 - 4.89 4.29Associated Acoustic ParametersFDA IEC UnitsPr.3 pra (MPa) 0.12W o P (mW) - 167.65 167.65 167.65min of [W .3(z1),ITA.3(z1)]Min. of [Pα(zs ),Ita.α(zs)](mW) -z1 zs (cm) -zbp zbp (cm) -zsp zb (cm) 1.50z@PII.3max z @ max. Ipi,α (cm) 2.70deq(zsp) deq(zb) (cm) 0.63fc fawf (MHz) 2.00 - 2.00 - 2.00 2.00Dim. of Aaprt X (cm) - 0.54 - 0.54 0.54Y (cm) - 1.38 - 1.38 1.38Other Information PD td (μsec)PRF prr (Hz)Pr@ PIImax Pr@ max. Ipi (MPa) 0.14deq@ PIImax deq@ max. Ipi (cm) 0.55Focal Length FLx (cm) - 9.10 - 9.10FLy (cm) - 4.50 - 4.50Ipa.3@ MImax Ipa.3@ MI max (W/cm2) 0.45OperatorControl Frequency (MHz) 2.00 - 2.00 - 2.00 2.00Focus (cm) 9.10 - 9.10 - 9.10 9.10Gate Size (cm) - - - - - -Sample Rate (Hz) - - - - - -a This Index is not relevant to this operating mode.b This transducer is not intended for transcranial or neonatal cephalic uses.c This formulation for TIS is less than that for an alternate formulation in this mode.# No data is provided for this operation condition since the maximum index value is not reported for the reason listed.