A p p e n d i x F A c o u s t i c O u t p u t R e f e r e n c eF - 2 6 I n s t r u c t i o n s f o r U s eAcoustic Output Reporting Table – Track 3, FDA 510(k) and IEC 60601-2-37(Per transducer/mode that exceeds MI or TI value of 1.0)Transducer model: 14L5 Operating mode: Color / PowerMI TIS TIB TICIndex Label Scan Non-scan Non-scanAaprt ≤1 Aaprt >1Maximum Value 1.45 2.27 - - - 2.34Associated Acoustic ParametersFDA IEC UnitsPr.3 pra (MPa) 3.66Wo P (mW) 30.66 - - 30.83min of [W.3(z1),ITA.3(z1)]Min. of [Pα(zs ),Ita.α(zs)](mW) -z1 zs (cm) -zbp zbp (cm) -zsp zb (cm) -z@PII .3max z @ max. Ipi,α (cm) 0.70d eq(zsp) deq(zb) (cm) -f c fawf (MHz) 6.35 6.36 - - - 6.36Dim. of A aprt X (cm) 0.12 - - - 0.12Y (cm) 0.3 - - - 0.30Other Information PD td (μsec) 0.61PRF prr (Hz) 7.57Pr @ PII max Pr @ max. Ipi (MPa) 4.26d eq@ PII max deq@ max. Ipi (cm) -Focal Length FLx (cm) 0.51 - - 0.51FLy (cm) 1.10 - - 1.10Ipa.3@ MImax Ipa.3@ MI max (W/cm2) 556.62OperatorControl Focus (cm) 1.00 0.51 - - - 0.51Res/Speed 5 3 - - - 3Ensemble Size 12 12 - - - 12Frequency (MHz) 6.25 6.25 - - - 6.25a This Index is not relevant to this operating mode.b This transducer is not intended for transcranial or neonatal cephalic uses.c This formulation for TIS is less than that for an alternate formulation in this mode.# No data is provided for this operation condition since the maximum index value is not reported for the reason listed.Acoustic Output Reporting Table – Track 3, FDA 510(k) and IEC 60601-2-37(Per transducer/mode that exceeds MI or TI value of 1.0)Transducer model: 14L5 Operating mode: Pulsed DopplerMI TIS TIB TICIndex Label Scan Non-scan Non-scanAaprt ≤1 Aaprt >1Maximum Value 1.22 - 2.91 - 2.91 2.22Associated Acoustic ParametersFDA IEC UnitsPr.3 pra (MPa) 3.09W o P (mW) - 82.07 82.07 89.11min of [W .3(z1),ITA.3(z1)]Min. of [Pα(zs ),Ita.α(zs)](mW) -z1 zs (cm) -zbp zbp (cm) -zsp zb (cm)z@PII.3max z @ max. Ipi,α (cm) 1.10deq(zsp) deq(zb) (cm)fc fawf (MHz) 6.32 - 7.43 - 7.43 6.34Dim. of Aaprt X (cm) - 2.63 - 2.63 2.63Y (cm) - 0.30 - 0.30 0.30Other Information PD td (μsec) 0.62PRF prr (Hz) 100.16Pr@ PIImax Pr@ max. Ipi (MPa) 3.93deq@ PIImax deq@ max. Ipi (cm)Focal Length FLx (cm) - 7.93 - 7.93FLy (cm) - 1.10 - 1.10Ipa.3@ MImax Ipa.3@ MI max (W/cm2) 435.76OperatorControl Frequency (MHz) 6.25 - 7.50 - 7.50 6.25Focus (cm) 1.50 - 7.93 - 7.93 7.93Gate Size (cm) 0.10 - 0.10 - 0.10 0.10Sample Rate (Hz) 100.16 - 24083 - 24083 24083a This Index is not relevant to this operating mode.b This transducer is not intended for transcranial or neonatal cephalic uses.c This formulation for TIS is less than that for an alternate formulation in this mode.# No data is provided for this operation condition since the maximum index value is not reported for the reason listed.