Chapter 5᤹5-305.2 ADJUSTAdjustment Mode5.2.1 ADJUST Table5.2.1.1 ADJUST᤹ Table 0004-7401CLC5100 / CLC4000 ADJUST > ADJXYT-5-17ADJUST > ADJXYSubitem DescriptionADJX Use it to make adjustments so that the image read start position (X direction,sub scanning direction) matches the reference point on the copyboard glass.Fineadjusts the distance from the original scanner home position sensor tothe read start position. Unit: Number of steps of the stepping motor. Follow the instructions on the next page.Settings range:200 to +200 (3.9 to +3.9mm)Unit:0.11 mm (approx.)ADJY Use it to make adjustments so that the image read start position (Y direction,main scanning direction) matches the reference point on the copyboardglass. Unit: Pixel Follow the instructions on the next page.Settings range:100 to +100 (2.5 to +2.5mm)Unit:0.06 mm (approx.)ADJS Use it to fineadjust the point at which the standard white plate is measuredfor shading correction data. Unit: Number of steps of the stepping motor. Scratches or dirt, if any, on the standard white plate can cause conspicuousvertical white lines on copies. If such is noted, shift the point of measurement using 'ADJS'.Settings range:0 to 45(0 to 4.8mm) Unit:0.11 mm (approx.)ADJJ Adjusts the preparatory time for the scanner motor.Settings range:450 to 550 Unit:0.1 msec (approx.)