4-28 CARESCAPE Monitor B850 2040384-004DLoad settingsNOTEThis option is available only when you log into Webmin remotely. The option isnot available locally at the patient monitor.1. If you have not already done so, “Log into Webmin” on page 4-4 and select theConfiguration tab.2. On the Configuration tab, select Settings.3. On the Sub-Modules for Settings menu, select Load.4. On the Load Settings window, enter the file name or click on Browse to select afile from the Choose file dialog box.5. Select Upload to load the settings.Activate settingsWhen the settings are activated, monitors configured for the U.S. will revert to theinternational factory default settings. To reload the U.S. factory default settings, usethe U.S. defaults CD and follow the procedure to “Load settings” on page 4-28.NOTEThe patient monitor needs to be in a discharged state.1. Log on to Webmin.2. Select the Configuration tab.3. On the Configuration tab, select Settings.4. On the Sub-Modules for Settings menu, select Activate.5. On the Activate Settings window, select the settings that you want to activate.6. Select Submit.The changes will take effect after the patient monitor is restarted.Software ManagementFor information, refer to the “CARESCAPE Monitor B850 Software InstallationInstructions”.