2040384-004D CARESCAPE Monitor B850 4-235. Select Submit.The change will take effect immediately.LicensingNOTEThis configuration applies only to the PDM.To activate the license for a PDM:1. If you have not already done so, “Log into Webmin” on page 4-4 and select theConfiguration tab.2. On the Configuration tab, select Modules.3. On the Sub-Modules for Modules menu, select Licensing.4. On the Licensing window, enter the activation code by the appropriate feature.5. Select Activate.NOTETo remove the license, select Remove next to the feature you want toremove.6. After adding and activating the license(s), perform the appropriate parametercheckout.The changes will take effect after the PDM is power-cycled.ECG filter configurationNOTEThis configuration applies only to the PDM.The ECG Filter is always enabled. It can be disabled temporarily, but it will alwaysdefault to an enabled state after a power cycle or reboot of the PDM.CAUTIONDo not disable the ECG Filter during clinical use.To set the acquisition ECG filter for the PDM:1. If you have not already done so, “Log into Webmin” on page 4-4 and select theConfiguration tab.2. On the Configuration tab, select Modules.3. On the Sub-Modules for Modules menu, select ECG Filter Configuration.4. On the ECG Filter Configuration window, select the applicable radio button toEnable or Disable the line frequency filter.5. Select Submit.The change will take effect immediately.