4-22 CARESCAPE Monitor B850 2040384-004D3. On the National Requirement window, select the applicable option:4. Select Submit.The national requirements changes take effect after the next system restart.Host asset settingsNOTEThe Host Serial Number field is view only and cannot be changed.To set the host asset number:1. If you have not already done so, “Log into Webmin” on page 4-4 and select theConfiguration tab.2. On the Configuration tab, select Host Asset Settings.3. On the Host Asset Settings window, enter the user-assigned host asset number inthe Change Value to column.4. Select Submit.The host asset changes take effect immediately.ModulesAsset settingsNOTE This configuration applies only to the PDM. The Device Serial Number field is view only and cannot be changed.To set the device asset number of a PDM:1. If you have not already done so, “Log into Webmin” on page 4-4 and select theConfiguration tab.2. On the Configuration tab, select Modules.3. On the Sub-Modules for Modules menu, select Assets Settings.4. On the Assets Settings window, enter the user-assigned asset number for thedevice in the Change Value to column.Value DescriptionNone Select the normal defaults.France Enable the following country specific monitoring: At least 1 audible and visual occurrence must remind eitheras an un-inhibitable audible and visual alarm or as anaudible and recurring recall repeated at least every 3minutes and also inhibitable. Heart Rate high alarm limit maximum 230 instead of 250. Reminder beep will sound every 2 minutes when alarmshave been silenced permanently.