6-4 CELL-DYN 3000 System Operator’s Manual9140240E — May 1995Calibration Chapter 6Complete instructions for using these criteria and a Calibration CriteriaChart are included with each procedure to facilitate the decision. Thechart is depicted in Table 6.1.Table 6.1: Calibration Criteria ChartThe calibration procedures have been divided into subsections thatconsist of a series of easy-to-follow steps. Always follow the entireprocedure unless specifically directed to skip to another section.Worksheets are provided at the end of each procedure to assist indetermining which parameters require calibration. For manualcalibration, worksheets may be used to assist in making the necessarycalculations. These worksheets may be duplicated as needed.NOTE: Always complete the Pre-Calibration procedures beforebeginning any calibration.Conventions Used in this ChapterA description of the Conventions used in this chapter is given in theintroductory section of this manual.Validation RangeCal Not RequiredCalibration RangeCal RequiredCalibration LimitDo Not CalWBC <1.5% >1.5% but <10% >10%RBC <1.0% >1.0% but <10% >10%HGB <1.0% >1.0% but <10% >10%MCV <1.0% >1.0% but <10% >10%PLT <3.0% >3.0% but <15% >15%