PaX-i3D Green (PHT-60CFO) User Manual202 Warnings and Precautions2.2 Electricity-related Safety Precautions■ Check the status of the power source, PC and cables prior to operating theequipment.■ Ensure that the on/off switch is set to off when the equipment is not in use.■ Always disconnect the power supply before cleaning the equipment.■ Always keep electrical cords away from hot appliances or radiators.■ DO NOT place the PC or peripheral equipment connected to the PC in theimmediate vicinity of the patient.■ The equipment and PC should be connected to a common protective earth.■ Never overload the equipment’s circuit by sharing it with too manyappliances.■ Use the same power circuit for the PC and the equipment.Combining this equipment with other devices■ Do not connect this equipment to devices which are not part of the system.■ Do not connect this equipment to a Multiple Portable Socket-Outlet (MPSO)or extension cord.Electromagnetic compatibility■ The PaX-i X-ray unit complies with IEC standard 60601-1-2.■ Medical electrical equipment is subject to special EMC preventivemeasures. It must be installed and operated as specified in the InstallationRequirements document.■ If high-voltage systems, radio link systems or MRI systems are locatedwithin 5 m of the unit, please observe the specifications stated in theinstallation requirements.■ Portable and mobile RF communications equipment may interfere withmedical electrical equipment. Therefore, the use of mobile wireless phonesin medical offices or hospital environments must be prohibited.