PaX-i3D Green (PHT-60CFO) User Manual926 Acquiring PANO ImagesAligning the Laser BeamsWARNINGBe careful not to project the laser beams directly into thepatient’s eyes as this could severely damage the patient’svision.A. Mid-sagittal plane laser beam:Position the Mid-sagittal plane laserbeam at the center of the patient’s face toprevent magnifications of the left or rightside in the final image.B. Frankfurt plane laser beam: TheFrankfurt plane laser beam should go fromthe top of the ear to the tip of the nosewhen the patient’s head is tilted back 10 –15 °.Frankfurt plane laserbeam UP buttonFrankfurt plane laserbeam DOWN buttonCompleting Patient PositioningA. After confirming patient positioning and beam alignment, adjust the templesupports to fit snugly on both sides of the patient’s head by clicking the templesupport adjustment button.B. Click the READY button after the patient has been properly positioned. NoX-ray will be emitted at this point.C. Proceed to section 6.4 Initiating X-ray Exposure.Frankfurt plane laserbeamMid-sagittal planelaser beam