PaX-i3D Green (PHT-60CFO) User Manual1208 Acquiring CT ImagesC. CT vertical laser beam: Position the CT vertical laser beam at the center ofFOV from the patient’s lateral.Completing Patient PositioningA. After confirming patient positioningand beam alignment, adjust the templesupports to fit snugly on both sides of thepatient’s head using the temple supportadjustment button.B. Click the READY button after the patient has been properly positioned. NoX-ray will be emitted at this point.If, for the FOV size of 50 x 50 or 80 X 50, the SCOUT image is to be acquired,click the SCOUT button.NOTESCOUT capture is featured in FOV size 50 X 50 and 80 X 50.In case of abnormal patient’s arch, the chinrest heightadjustment through the SCOUT feature enables the userto acquire better image. For details about imaging in theSCOUT mode, refer to the section “8.3 Acquiring a ScoutImage”.C. Proceed to the section 8.4 Initiating X-ray Exposure. But if the SCOUTfeature is selected, continue to the next section 8.3 Acquiring a SCOUTImage, to the section 8.4 Initiating X-ray Exposure.