FD Series EN-203.8 Legal for Trade SettingsThe FD Series is designed to comply with OIML, EEC, NTEP and Measurement Canada weights andmeasures regulations. These approvals may be pending; contact Ohaus for further details aboutavailability.Note:Consult your local Weights and measures office for their regulations before placing the scale into aLegal for Trade application.When the LFT menu item setting is On, the following conditions apply:• The Calibration (CAL) sub-menu is hidden.• The Calibration Lock (LCAL) menu item is hidden.• The Auto Zero Tracking (AZt) menu item is set to 0.5 d.• lb:oz unit is disabled.When the weights and measures official has approved the scale for use, it should be sealed using oneof the methods shown in Section 3.9.3.9 Sealing the ScaleWhen used in conjunction with the Lockout menu and the security switch, the scale may be sealedto prevent or detect unauthorized changes to scale settings. For Legal for Trade applications, setthe LFT menu item to ON, set the security switch to the On position, then seal the scale to preventaccess to the metrological parameters. Seal the scale according to the local weights and measuresregulations.Sliding the switch to the ON position in FDxxH models will freeze the menu item settings in theircurrent stateTo regain access to the scale settings, break the seal and set the security switch to the OFF position.On OffFigure 3-5. Sealing methods.Security SwitchFigure 3-4. Security Switch Settings.Wire Seal Paper SealWire SealRemove the spacers under the cross drilledscrews. Feed the wire through the crossdrilled holes.Paper SealPlace the paper seals over the screws andsealing spacers.Visit www.balances.com to see the full line of Ohaus BalancesBuy your next Ohaus Balance online at https://scaleman.com/