FD Series EN-14Span Calibration Procedure• When CAL is displayed, press Yes to enter the calibrationsub-menu.• When Span is displayed, press Yes to begin the Spancalibration process.• Clear the pan if prompted to do so, then press Yes.....• The display will show “--C--”, followed by the spancalibration weight value.• If desired, press No to change to an alternate spancalibration weight value.• Place the indicated calibration weight on the scale andpress Yes.• The display will show “--C--”, then “done” and thenrevert to weighing. Remove the calibration weight.Note: The calibration procedure can be canceled at anytime by pressing the Exit button or turning the scale off.Linearity Calibration Procedure• When LIN is displayed, press Yes to begin the linearitycalibration process.• Clear the pan if prompted to do so, then press Yes.• The display will show “--C--”, followed by the firstcalibration weight value.• If desired, press No to change to an alternate firstcalibration weight value.• Place the indicated calibration weight on the scaleand press Yes.• The display will show “--C--”, followed by the secondcalibration weight value.• Place the indicated calibration weight on the scaleand press Yes.• The display will show “--C--”, then “done” and thenrevert to weighing. Remove the calibration weight.Note: The calibration procedure can be canceled at anytime by pressing the Exit button or turning the scale off.End• Press Yes to leave this sub-menu and go to the next.• Press No to return to the top menu item in this sub-menu.kkkkkVisit www.balances.com to see the full line of Ohaus BalancesBuy your next Ohaus Balance online at https://scaleman.com/