Language Dictionary 3Keysight Models 6811C, 6812C, and 6813C Programming Guide 203TRIGger:SEQuence2:PHASeTRIGger:SYNChronize:PHASeThese commands set the phase angle with respect to an internal phase referencewhen the TRIGger:SYNChronize:SOURce is set to PHASe. The range is from ā360to +360 degrees.TRIGger:SEQuence3TRIGger:ACQuireWhen the trigger subsystem has been initiated, these commands generate ameasurement trigger regardless of the selected trigger source. The measurementtrigger causes the ac source to digitize the instantaneous output voltage andcurrent for several output cycles and store the results in a buffer.The FETCh commands return the requested calculation from this acquired data.When the measurement completes, the WTG bit in the Status OperationCondition register is cleared.Command syntax TRIGger:SEQuence2:PHASe TRIGger:SYNChronize:PHASe Parameters ā360 to +360 (degrees) | MAXimum | MINimum*RST value 0Examples TRIG:SYNC:PHAS 90 TRIG:SEQ2:PHAS 180Query syntax TRIGger:SEQuence2:PHASe?TRIGger:SYNChronize:PHASe?Returned parameters Related commands ABOR TRIG TRIG:DEL TRIG:SYNC INIT INIT:CONT*TRG WAICommand syntax TRIGger:SEQuence3[:IMMediate]TRIGger:ACQuire:[:IMMediate]Parameters NoneExamples TRIG:ACQ TRIG:SEQ3:IMMRelated commands ABOR TRIG TRIG:DEL TRIG:SYNC TRIG:SYNC:PHASINIT INIT:CONT *TRG WAI