Language Dictionary3Keysight Models 6811C, 6812C, and 6813C Programming Guide141LIST:SHAPeThis command sets the sequence of the waveform shape entries. The order inwhich the shapes are given determines the sequence in which the list of shape willbe output when a list is triggered. Changing list data while a subsystem is in listmode generates an implied ABORt. The following shapes may be specified:–SINusoid: A sine wave is output–SQUare: A square wave is output–CSINusoid: The output is a clipped sine wave. Both positive and negative peakamplitudes are clipped at a value determined by the FUNCtion:CSINusoidcommand.–
: The output shape is described by one of the user-defined waveformtables.The maximum peak voltage that the ac source can output is 425 Vpeak. Thisincludes any combination of voltage, voltage offset, and function shape values.Therefore, the maximum value that can be programmed depends on thepeak-to-rms ratio of the selected waveform. For a sine wave, the maximumvoltage that can be programmed is 300 Vrms.LIST:SHAPe:POINts?This query returns the number of points specified in LIST:SHAP. Note that itreturns only the total number of points, not the point values.Command syntax[SOURce:]LIST[:SHAPe] {, }ParametersSINusoid | SQUare | CSINusoid |