3 Language Dictionary104 Keysight Models 6811C, 6812C, and 6813C Programming GuideOUTPutThis command enables or disables the ac source output. The state of a disabledoutput is an output voltage amplitude set to 0 volts, with output relays opened.The query form returns the output state.For Keysight 6812C/6813C Option 026 only, this command controls the on/offstate of both the main and auxiliary outputs.OUTPut:COUPlingThis command enables AC or DC output coupling. When the output coupling isset to AC, a DC leveling loop attempts to maintain zero average output voltage.The loop has a corner frequency of about 2 Hz. It will not prevent short transientwaveforms that may have non-zero average voltage, but will cause a settlingtransient to an average value of 0 volts.The output coupling must be set to DC to obtain DC output with VOLTage:OFFSet,or to generate output transients that have net DC components.Command syntax OUTPut[:STATe] Parameters 0 | 1 | OFF | ON*RST value OFFExamples OUTP 1 OUTP:STAT ONQuery syntax OUTPut[:STATe]?Returned parameters 0 | 1Related commands *RCL *SAVCommand syntax OUTPut:COUPling Parameters AC | DC*RST value ACExamples OUTP:COUP DCQuery syntax OUTPut:COUPling?Returned parameters Related commands *RCL *SAV