CONFigure[:PRIMary]:CURRent[:DC]SyntaxCONFigure[:PRIMary]:CURRent[:DC] [{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]DescriptionFirst, this command resets all (primary) DC current measurement parameters and trigger parameters totheir default values. Then, it configures the instrument for DC current measurements but does not initiatethe measurement.The CONFigure command does not place the instrument in the "wait-for-trigger" state. Use the INITiate orREAD? command in conjunction with CONFigure to place the instrument in the "wait-for-trigger" state.ParametersName Type Range of Values Default Value<range> Numeric Desired range in Amps— 100 μA (MIN)— 1 mA— 10 mA— 100 mA (DEF)— 1 A— 10 A (MAX)— Auto range (AUTO)Auto range (AUTO)<resolution> Numeric — 3.00E-5— 2.00E-5— 1.50E-61.50E-6 (5½ digits)Remarks— You can allow the instrument to automatically select the measurement range using auto ranging oryou can select a fixed range using manual ranging. Auto ranging is convenient because the instrumentdecides which range to use for each measurement based on the input signal. For faster measurements,use manual ranging on each measurement (some additional time is required for auto ranging since theinstrument has to make a range selection).— If you select auto ranging (by specifying "AUTO" or "DEF"), an error will be generated if you specify a dis-crete value for the parameter. When auto ranging is combined with a discrete resolution,the instrument cannot accurately determine the integration time (especially if the input signal isKeysight 34450A Programmer's Reference 672 SCPI Programming