MEMory:STATe:RECall:AUTOSyntaxMEMory:STATe:RECall:AUTO de>MEMory:STATe:RECall:AUTO?DescriptionThis command disables or enables the automatic recall of the power-down state (state 0) when power isturned on. Select "ON" to automatically recall the power-down state (location 0) when power is turned on.Select "OFF" to issue a Factory Reset (and state 0 is not automatically recalled) when power is turned on.ParametersName Type Range of Values Default Value<mode> Boolean {OFF|0|ON|1} This is a required parameterRemarks— With the *SAV command, you can use location 0 to store a fifth instrument state. However, keep inmind that location 0 is automatically overwritten with the power-down state when power is cycled.— When shipped from the factory, storage locations 1 through 5 are empty and the instrument is con-figured to NOT automatically recall the power-down state (location 0) when power is restored.— A Factory Reset (*RST command) does not affect the value of <mode>.Return FormatThe query command returns "0" (OFF) or "1" (ON).ExamplesThe following command enables automatic recall of the power-down state.MEM:STAT:REC:AUTO ONThe following query returns the automatic recall setting.MEM:STAT:REC:AUTO?Typical Response: 02 SCPI Programming164 Keysight 34450A Programmer's Reference