STATus:QUEStionable:CONDition?SyntaxSTATus:QUEStionable:CONDition?DescriptionThis command queries the condition register for the Questionable Data Register group. This is a read-only register and the bits are not cleared when you read the register.For more information on the SCPI Status System for the Keysight 34450A, see the Status SystemDiagram.Remarks— The condition register bits reflect the current condition. If a condition goes away, the correspondingbit is cleared in the condition register.— The following table lists the bit definitions for the condition register in the Questionable Data group.Bit Number Decimal Value Definition0 Voltage Overload 1 Only reported as event. Returns "0".Read the Event Register.1 Current Overload 2 Only reported as event. Returns "0".Read the Event Register.2-3 Not Used Not Used "0" is returned.4 Temperature Overload 16 Only reported as event. Returns "0".Read the Event Register.5 FrequencyOverload/Underflow32 Only reported as event. Returns "0".Read the Event Register.6-7 Not Used Not Used "0" is returned.8 Calibration Corrupt 256 At least one calibration constant is corrupt.9 Resistance Overload 512 Only reported as event. Returns "0".Read the Event Register.10 CapacitanceOverload/Underflow1024 Only reported as event. Returns "0".Read the Event Register.11 Lower Limit Failed 2048 Reading is less than lower limit in limit test.12 Upper Limit Failed 4096 Reading is greater than upper limit in limit test.13-15 Not used Not Used "0" is returned.— A Factory Reset (*RST command) will clear all bits in the condition register.Keysight 34450A Programmer's Reference 2792 SCPI Programming