PRN 50M+Cleaning points to note● Warranty does not cover any damages caused by using substances or methods.● Manufacturer makes no claims as to the efficacy of the listed substances ormethods to control infection.● Always dilute cleaning and disinfectant agents according to their manufacturer’sinstructions. Always consider your hospital guidelines as well.● Your hospital guidelines permitting, all cleaning and disinfection activities canbe carried out at the bedside.● Do not let liquid pool around connection pins. If this should happen, blot dry with asoft, lint-free cloth.● Do not use excessive drying techniques, such as oven, forced heat, or sun drying.● Never connect any device until it is thoroughly dry.● If you discover any signs of deterioration or damage in the device, discontinue itsuse.General cleaning instructionsFollow these instructions to clean the device.1. Turn off the power to the device.2. Disconnect the device from the power supply.3. Remove all cables.4. Dampen a soft lint-free cloth with one of the permitted detergents.5. Wring excess liquid from the cloth and wipe the exterior surface. Pay specialattention to hard-to-clean areas, like grooves and crevices.● Remove any soil as soon as possible after use by wiping the device.● Any contact of cleaning solutions with metal parts may cause corrosion.● Do not damage or bend connector pins when cleaning or drying.● Do not let fluid pool around connection pins. If this happens, blot dry witha cotton swab or soft cloth.6. Wipe off the cleaning solutions with a clean, lightly moistened cloth.7. Visually ensure that the device is clean. If visible soil remains, repeat cleaningprocedure until the device is thoroughly clean.8. Dry thoroughly with a dry, lint-free cloth and let air dry for at least 30 minutes.Drying times may vary based on the environmental conditions.9. Reconnect the device to the power supply.10. Reconnect all cables.11. Turn on the power to the device.Cleaning internal componentsOn a regular basis, qualified service personnel should open the writer and blow dustfrom the unit with compressed air.66 Cleaning and disinfecting 2110145-001