CARESCAPE B650 MonitorWARNING Do not autoclave any part of the system with steam (includingcables or leadwires) or sterilize with ethylene oxide orradiation.WARNING Do not pour or spray any liquid so that it may seep intoconnections or openings.WARNING Never use conductive solutions, oxidizing compounds, wax, orwax compounds to clean devices, cables or leadwires.Cleaning cautionsCAUTION Do not apply pressurized air or gas to any outlet or tubingconnected to the monitor. Pressure may destroy sensitiveelements.Permitted detergents● Water● Ethanol 70 to 96% by volume● Isopropyl alcohol max. 60% by weight● Enzymatic or neutral pH detergentsCleaning points to note● Warranty does not cover any damages caused by using other than GE approvedsubstances and methods.● GE makes no claims as to the efficacy of the listed substances or methods tocontrol infection.● Your hospital guidelines permitting, all cleaning and disinfection activities canbe carried out at the bedside.● Do not let liquid pool around connection pins. If this should happen, blot dry with asoft, lint-free cloth.● Do not use excessive drying techniques, such as oven, forced heat, or sun drying.● Do not spray cleaner directly on the display screen.● Never connect any device or applied part to a patient until it is thoroughly dry.● If you discover any signs of deterioration or damage in the device, discontinue itsuse.Setting the touchscreen off for cleaningYou can set the touchscreen feature off for 30 seconds at a time when you need toclean the screen. The countdown timer appears on the screen.1. Select Monitor Setup > Main Setup.2. Select Touchscreen Off.3. You can enable the touchscreen immediately by pressing any monitor hardkey orthe Trim Knob, or by selecting Cancel in the Touchscreen Off display.50 Cleaning and disinfecting 2110145-001