PaX-i3D Green (PHT-60CFO) User Manual148Appendices1 Recommended X-ray Exposure TableA. PANO[Standard / TMJ / Sinus / CARPUS]Gender / Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)Man 75 / 14 74 / 14 73 / 14Woman 74 / 14 73 / 14 72 / 14Child 68 / 10 67 / 10 66 / 10[Fast Mode]Gender / Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)Man 75 / 16 74 / 16 73 / 16Woman 74 / 16 73 / 16 72 / 16Child 68 / 16 67 / 16 66 / 16* Tube voltage and current variation range in the PANORAMIC mode: 50 ~ 80 kVp, 4 ~ 16 mAB. CEPH[Lateral]Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)One shot Scan FastScan One shot Scan FastScan One shot Scan FastScanMan 92/14 87/10 97/14 90/14 85/10 95/14 88/14 83/10 93/14Woman 90/14 85/10 95/14 88/14 83/10 93/14 86/14 81/10 91/14Child 86/10 82/10 87/12 84/10 80/10 85/12 82/10 78/10 83/12[PA / SMV / Waters View]Figure Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)One shot Scan FastScan One shot Scan FastScan One shot Scan FastScanMan 94/14 90/10 99/14 92/14 88/10 97/14 90/14 86/10 95/14Woman 92/14 88/10 97/14 90/14 86/10 95/14 88/14 84/10 93/14Child 90/10 85/10 89/12 88/10 83/10 87/12 86/10 80/10 85/12GenderSensorTypeGenderSensorType