PaX-i3D Green (PHT-60CFO) User Manual1077 Acquiring CEPH ImagesCAUTIONAsk the patient not to block the ruler on the nasalpositioner by his / her fingers. It could deteriorate imagequality.C. Click the READY button after the patient has been properly positioned. NoX-ray will be emitted at this point.D. Proceed to section 7.3 Initiating X-ray exposure.7.3 Initiating X-ray ExposureThe X-ray exposure method and process is the same for all CEPH modes. The exampleand images below are taken from an X-ray performed in CEPH Lateral mode.WARNING- Do not operate the PC during exposure. Failure tocomply with this instruction may cause the system tomalfunction.- The operator shall observe the X-ray safety regulationsapplicable to their area at all times during the operationof this equipment.CAUTIONIf an emergency occurs during image acquisition, releasethe exposure switch to cease X-ray emission.A. Leave the X-ray room and close the door. The operator must visually monitorthe patient at all times during image acquisition.B. Press and hold down the exposure switch until image acquisition is complete.During X-ray irradiation, ensure that- The LED lamp on top of the equipmentturns orange to indicate X-ray emission.- The warning lamp outside the X-rayroom turns on.- The sound (beep or music: optional) goes off.- The radiation symbol on the upper leftcorner of GUI turns yellow to indicate X-Rays are being emitted.X-ray ONExposure indicatorlight turns orangeOrange: Exposure