- 118 -34.1.1. Module technical specificationMaximal number of convertor units 8 388 608OIML class IIINumber of verification units 6000eMaximal signal gain 19mVMaximal voltage per verification unit 3,3 μVMinimal voltage per verification unit 1μVMinimal strain gauge impedance 90ΩMaximal strain gauge impedance 1200ΩStrain gauge converter supply voltage 5VStrain gauge converters connection 4 or 6 wires + shield34.1.2. Weighing platform signal wires coloursStrain gaugesensordeterminersColoursaccording to theRADWAGstandardW eighing moduledeterminers+INPUT brown +5V-INPUT green AGND+OUTPUT yellow +IN- OUTPUT white -IN+SENSE gray +REF- SENSE pink - REFSCREEN yellow-green (according to the ruleof connecting shields)