78ALARMS, SAFETY, AND SELF-DIAGNOSISAlarm or safetyfunction Conditions Display Buzzer RemotealarmSafetyoperationUV lamp failure(MCO-170AICUVHL/MCO-170AICUVL, or when an optional MCO-170UVS isinstalled)The UV lamp burns out“Err18: UV Lamp Abnormal” is displayedin the message display field.Intermittenttone ON -----New UV lampreplacement(MCO-170AICUVHL/MCO-170AICUVL, or when an optional MCO-170UVS isinstalled)The accumulated ON time reachesapprox. 5,000 h“Warning: UV Bulb Life” is displayed inthe message display field. ----- ----- -----CommunicationerrorWhen communication between LCDtouch panel and control substrate isdied out or unstable.“Err56: Communication Failure” isdisplayed in the message display field. ----- ----- -----●Table 5~7 show the behavior of the alarm (buzzer) and Ring Back function when pressing Buzzer key.Table 5 In the cases of other than table 6 or table 7.Remote Alarm setting Ring BacksettingBuzzer from CO 2 incubator Remote AlarmWhen pressingthe Buzzer keyWhen the Ring Backset time passesWhen pressingthe Buzzer keyWhen the Ring Backset time passesON: Non-interlockwith Buzzer keyON OFF(Alarm is notcanceled)ON ON ON(Under continuation)OFF OFFOFF: Interlockwith Buzzer keyON ON OFF (Alarm isnot canceled)ONOFF OFF OFFNote: Resolve the cause of the alarm in reference to page 77-79 because the alarm itself is notdeactivated by pressing Buzzer keyTable 6 In the cases of high limit temperature alarm or Err38 (The outer door opens during H2O 2decontamination; refer to next page)Remote Alarm setting Ring BacksettingBuzzer from CO 2 incubator Remote AlarmWhen pressingthe Buzzer keyWhen the Ring Backset time passesWhen pressingthe Buzzer keyWhen the Ring Backset time passesON: Non-interlockwith Buzzer keyONON ON(Under continuation)ON(Continue)ON(Under continuation)OFFOFF: Interlockwith Buzzer keyONOFFNote: Close the outer door when Err38 is activated.Table 7 In the cases of Err01 (CO2 gas cylinder empty), Err11, 12 (CO 2 sensor error), Err18 (UVlamp failure) or door alarm*.Remote Alarm setting Ring BacksettingBuzzer from CO 2 incubator Remote Alarm settingWhen pressingthe Buzzer keyWhen the Ring Backset time passesWhen pressingthe Buzzer keyWhen the Ring Backset time passesON: Non-interlockwith Buzzer keyON OFF(Alarm iscanceled)OFF(Alarm is alreadycanceled)OFF(Alarm iscanceled* 2)OFF(Alarm is alreadycanceled* 2)OFFOFF: Interlockwith Buzzer keyONOFF*2: In the door alarm, the remote alarm does not work.Note: Err01 recovery is achieved automatically by correctly connecting the new CO 2 gas cylinder andpressing the Buzzer key to stop the buzzer. In addition, when the optional MCO-21GC is installed and thegas supply is switched to the reserve gas cylinder, press the Buzzer key and replace the gas cylinder. |
This manual is suitable for:
Product
Menu
Learn More
Menu
Legal
Menu
Copyright © 2024 OneStart. All Rights Reserved