272 Keysight E1420B Universal Counter User GuideNew Capabilities SCPI CommandsSCPI CommandsSyntaxThe following phase measurement commands have been added::CONFigure[1][:VOLTage]:PHASe [[,]]:MEASure[1][:VOLTage]:PHASe? [[,]]:READ[1][:PHASe]?[:SENSe[1]]:FUNCtion "[VOLTage:]PHASe"The syntax for the phase measurement commands is shown above. While the and parameters are accepted, they are ignored.Furthermore, the optional numeric suffix of 1 implies a Phase 1 relative to 2measurement.Command DescriptionsThe :CONFigure, :MEASure?, and [:SENSe]:FUNCtion commands reconfigureseveral instrument settings to optimize the measurement of Phase 1 relative to 2.The :CONFigure and :MEASure? commands invoke identical settings to set ageneric configuration for this measurement. For those applications that requirefiner control of the instrument configuration, the [:SENSe]:FUNCtion commandpresets fewer settings. This allows you to set the instrument to a specificconfiguration using other [:SENSe] commands.Table F-1lists the instrument settings that are initialized whenever a phasemeasurement is selected. Bold type indicates mandatory settings. Attempting tochange these settings generates the -221, Settings conflict error message. Theother settings can be changed to customize the measurement for a specificapplication.