Keysight E1420B Universal Counter User Guide 175MEMORY Subsystem (Option 040) E1420B Command ReferenceExampleSetting the VME memory size:MEM:VME:SIZE 64000 sets memory size to 64 kbytesComments– Memory size can be specified in decimal or Hexadecimal (#H...).– The MIN parameter sets the memory size to 0 bytes and MAX sets thememory size to 12582912 bytes (#HCOOOOO).– The Address and size are interdependent and setting one may affect theother.– *RST condition: VME:MEM:SIZE 0.E1420B Command Reference 5-51:VME:SIZE?MEMory:VME:SIZE? [MINimum | MAXimum] returns one of the following numbersto the output buffer:– The present memory size (in decimal) selected if no parameters arespecified.– The smallest memory size available (0) if MIN is specified.– The largest memory size available (12582912) if MAX is specified.ExampleQuerying the VME memory size:MEM:VME:SIZE 64000 Set memory size to 64 KbytesMEM:VME:SIZE? Query counter to return allocated VMEmemory size."ENTER" statement. Enter the value into computer:VME:STATeMEMory:VME:STATe enables or disables the use of external VMEmemory for data storage.ParametersExampleEnabling VME memory:CONF:FREQ DEF,DEF Function: Setup for a 10Mhz frequencymeasurement.Parameter Name Parameter Type Range of Values Default Valuemode Boolean OFF,0|ON,1 OFF, 0