16 Keysight M9195A PXIe Digital Stimulus/Response Service GuideUsing Keysight N7876A Calibration and Adjustment SoftwareTest SetupSelf Test is run without any connections to the M9195A front panel. The otherPerformance Verification tests require only one setup which includes a high-enddigital multimeter (3458A) with multifunction switch unit (34980A) holding twomicrowave switch (34947A) and an RF multiplexer (34941A) modules. A specificcable harness (Y1254A or Y1255A) is required. The digital multimeter connectsto the host computer or embedded controller through a GPIB cable.For more details, refer to the help system for N7876A PCI Digital IO Calibration.Equipment RequiredTest Equipment Recommended ModelDigital multimeter 3458AMultifunction switch unit 34980AQuad 1x4 RF multiplexer• 2 required34941ATriple 1x2 SPDT microwave switch 34947A Option 004SMA breakout cable Y1255A, 2 metersCable, flexible 3.5 m (m) 3.5 mm (m), 9-inch• 6 requiredHubner & Suhner 5063-1531Cable, flexible 3.5 m (m) 3.5 mm (m),12-inch• 2 requiredHubner & Suhner 5063-1532Cable, BNC (m) to BNC (m) 10503AAdapter, BNC (f) to SMA (m) 1250-1200Adapter, BNC (f) to Dual Banana 1251-2277PXIe chassis M901xXEmbedded controller for PCIe chassis (or below) M903xXHost PC with GPIB interface GPIB interface: Keysight 82351B orNI GPIB-USB-HS.GPIB cable, 1 meter