4 Keysight M9195A PXIe Digital Stimulus/Response Service GuideHigh-Level Diagnostic Tools, Processes and References– Solid Blue indicates that conditions for Solid Green are met and that theIVI driver for the M9195A is now initialized.If the status LED does not go from solid green to solid blue after the SoftFront Panel comes up and connects with the M9195A, there may be anLED problem. If the Soft Front Panel does not come up, see “CheckSoftware Installation” on page 5.If the LED goes solid blue and the Soft Front Panel comes up, check thatthe drivers are loaded correctly by flashing the blue LED from the SFP.See “Check Module Status LED” on page 6.If an incorrect LED condition repeats and continues, go to “Returning aModule for Service” on page 22.2 Does the M9195A appear correctly on Keysight Connection Expert?– Bring up Connection Expert and click on the “PXIe/AXIe Chassis” tab thencheck to see if the M9195A in question appears in the correct slot.If the module is not visible, try shutting down the system, cycling power,and re-starting.If the M9195A still does not show up in Connection Expert then powerdown the chassis and repeat the process with the M9195A installed inanother slot.If the M9195A does not show up in Connection Expert, but the Soft FrontPanel (SFP) connects to it, then “Check Software Installation” and, ifnecessary, re-install all required software in the order describe in theM9195A Startup Guide.If the Module continues to not show up in Connection Expert, go to“Returning a Module for Service” on page 22.3 Does the M9195A (SFP) connect?If the Connection Expert as well as the SFP does not connect, go throughStep 2 above.– If the SFP fails to connect to the M9195A, then “Check SoftwareInstallation” on page 5 and, if that is OK, try repeating the process with theM9195A installed in another slot.Re-install all required software in the order describe in the M9195A StartupGuide.– If the SFP connects, go to “Troubleshooting with the Soft Front Panel” onpage 6.