KM9106 Operators Manual64. NORMAL START UP SEQUENCE4.1 Every Time You Use The AnalyserBEFORE SWITCH-ON CHECK THAT:the Oxygen sensor is connectedthe particle filter is not dirtythe sulphur filter is fitted for heavy oil or coalthe water trap and probe line are empty of waterall hose connections, etc, are properly madethe paper roll is fittedthe probe is sampling ambient FRESH airthe water trap is verticalthe flue temperature is connectedthe instrument is placed on a clean, flat, level surfaceSwitch ON the instrument by pressing4.2 Automatic CalibrationDuring this sequence the analyser pumps fresh air into the sensors to allow toxicsensors to be set to zero and the Oxygen sensor to be set to 20.9 %.After switch-on the analyser willbriefly display the Kane logo andtelephone numbers:-777 KANE-MAY 777TEL +44 (0) 1707 375550FAX +44 (0) 1707 393277Followed by the logger menu screen :- 9 9 SELECT FUNCTION 9 9. . LOGGER CONTROL2 . . QUINTOX CONTROLUse the and keys toposition the cursor over 2 . .QuintoxControl.Press to access QuintoxControl9 9 SELECT FUNCTION 9 91 . . LOGGER CONTROL. . QUINTOX CONTROL1ON2ENTER