Fluid Volumes9-6 COBE ® Spectra™ Apheresis System • Essentials GuideNet Additional Saline Returned to the Donor orPatientUse Table 9-3 to determine the net amount of additional saline returned toa donor/patient at the end of Rinseback mode. .Table 9-3: Net amount of additional saline returned to donor/patientProcedure Total Volume of Saline ProcessedDuring RinsebackAmount of Saline Diverted toWaste Bag During Waste DivertNet Additional Saline Returned toDonor/PatientELPPlatelet Depletion+310 mL –120 mL +190 mLLRS and LRST +320 mL –120 mL +200 mLLPEMNC and PMN CollectionWBC Depletion (MNC and PMN)+413 mL –150 mL +263 mLAutoPBSC +335 mL –150 mL +185 mLTPE +345 mL –150 mL +195 mLRBCX* +345 mL –100 mL +245 mL* Rinseback is not recommended for RBCX procedures.