Appendix E: Installing Wall Mount UnitsWall Mount Bracket InstallationService Manual E-3E.2 Wall Mount Bracket InstallationAfter the wall is finished and the room is completed you are ready to install the wall mount bracket.1. Utilizing a level, position the wall mount bracket on the wall and confirm that the bracket is bothhorizontally and vertically leveled. The bracket shall be positioned on the wall such that the four bracketholes be placed on the center of the wall tracks and the top edge of the wall mount bracket must be60.8” +/- 0.1” (154.4 +/- 0.25 cm) from the floor for installations to be used with bassinets, and 24.8" +/-0.1" (63 cm +/- 0.25 cm) from the fixed surface for installations to be used with fixed surface (SeeFigure E-2).2. Use the wall mount bracket as a template to mark the mounting holes on the wall.Warning: Always follow the instructed specifications to install the bracket. Incorrectinstallation will result in improper operation and may adversely affect the patient.FIGURE E-2. Positioning of Wall Mount Bracket