49Display refresh rate ( 5pd )The periodic time to refresh the display. This parameter influences "Baud rate", "Data output pause"and the data output rate of "Stream mode".Decimal point ( pnt )The decimal point format can be selected.Auto display-ON ( p-on )When the AC adapter is plugged in, the display is automatically turned on without the ON:OFFkey operation, to display the weighing mode. Used when the balance is built into an automated system.One hour warm up is necessary for accurate weighing.Auto power-OFF(p-off)This is a function to turn off only the display automatically when there is no operation made for acertain amount of time (approximately 10 minutes) while the power is on.Minimum display( )When weighing with rough precision, the minimum display can be turned off without keyoperation. This is useful when built into an automated system/Buzzer(beep)Select ON/OFF for the built-in buzzer that sounds when a key is operated or the status changes.Tare value record( )After turning on the power supply, the display will not be automatically set to zero, and it will startfrom the previous weighing value. This is useful when a hopper, etc. is attached to the weighing panand the power needs to be turned off while weighing discharge.Backlight brightness( )Select the brightness of the backlight of the LCD display.Bubble sprit level lightning( )Select ON/OFF for the LED that illuminates the bubble sprit level.