2000-OSM, F1 6-19CAUTIONSee "AIR PURGING" in Section 4 for the override function. Do not performoverride until you have read Section 4 completely and you understand and canperform the procedure requirements.Install a temperature monitor and verify ambient temperature. Ensure air flow into the Controllerhousing.Set Point UnacceptableSymptom: Alarm Condition.Any one of the following alarm messages or conditions could be present:On Background Screen ***ALARM:*** (flashing)On Manual Temperature Zone (#) (flashing)Control Mode Screen *Temp Bd Alarm* (flashing)On Alarm Screen Temperature Board:*Zone Specific*** TEMP ZONE (#) Alarm ****Set Point FailedTo modify the temperature configuration table, refer to the "DIGITAL TEMPERATURECONTROLLER" in Section 4.Reconfiguration FailureSymptom: Alarm Condition.Any one of the following alarm messages or conditions could be present:On Background Screen ***ALARM:*** (flashing)On Manual Temperature Zone (#) (flashing)Control Mode Screen *Temp Bd Alarm* (flashing)On Alarm Screen Temperature Board:*Zone Specific*** TEMP ZONE (#) Alarm ****Reconfiguration FailedRemedy: Perform the temperature board reconfiguration procedure. Be certain to hold down theswitch and respond by pressing “Y”. To modify the temperature configuration table, refer to the"DIGITAL TEMPERATURE CONTROLLER" in Section 4.Out of ControlCAUTIONIt is the customer’s responsibility to ensure that the area is safe and and hazard-free, and will remain so the entire time the analyzer is open. This responsibilityincludes ensuring adequate ventilation in analyzer shelter and obtaining properwork permits, etc.