Endura AZ30 series integral probe and remote transmitterCombustion oxygen monitor 6 ProgrammingCOI/AZ30E–EN Rev. D 656.4.8 CommunicationUsed to configure HART communication parameters andcyclic output communication via the IrDA interface on the frontof the transmitter.Parameter Comment / [Range] DefaultHARTHART Poll Address Used to enter a unique device address.When set to 0, Current Output 1 is active and providesan analog output proportional to its source. HARToperates in point-to-point mode.If set between 1 and 15, Current Output 1 is fixed at 4mA and HART operates in multi-drop mode.Note. See Section 8.2, page 74 for multi-dropconfiguration details.[0 to 15]0HART Device ID The unique ID for the HART device. Factory-setread-only IDHART Tag Used to enter a user-defined tag for the HART device(transmitter), selected from the alphanumeric characterset displayed when the Edit button is selected at thissub-parameter.[0 to 8 characters]N / AHART Device Message Used to specify a a user-defined message for the HARTdevice (transmitter), selected from the character setdisplayed when the Edit button is selected at thissub-parameter.[0 to 16 characters]N / AHART Descriptor Used to specify a unique HART descriptor, selected fromthe character set displayed when the Edit button isselected at this sub-parameter.Maximum of 32 characters.N / ADevice Install Date The time and date of installation. User-setFinal Assembly No. The device's (transmitter’s) final assembly (ID) number. Factory-setTable 6.13 Communication MenuMenuExit SelectCommunication