4.CALIBRATION204. CALIBRATION"Calibration" refers to an operation whereby matching between the F340A and a straingauge sensor is obtained. The F340A uses the two calibration methods as describedbelow.◇ Equivalent Input CalibrationThis approach uses no actual loads but key entry of the rated output value of the straingauge sensor (mV/V) and the rating value (value to be displayed).This method is simple and employed when actual loads cannot be applied.For exampleGain will be automatically decided by registering the values indicated as follows:for load: 2.001mV/V - 100.0kgffor pressure: 2.002mV/V - 10.00kgf/cm 2, andfor torque: 2.502mV/V - 15.00kgf•m.F340AStrain gauge sensorRated output value(mV/V)+Indicated valueA data sheet will be attached to a strain gauge sensor you buy.On the data seat, the following values are listed.Capacity ................................... Load (unit: kg, t, etc.)Rated Output............................ Voltage (unit: mV/V)Non-Linearity, Hysterisis,Input Resistance,Output Resistance,Zero Balance, etc.The Capacity and the Rated Output are necessary values for theequivalent input calibration. Input these two values to F340A.