5.4 Wiring5.4.1 Terminal Treatment of CablesFollow the procedures below to treat the terminals (at the converter side) of various cables and install thecables to the terminal block. Use appropriate cables based on the description in Section 5.1 "Cables."Crimp a round type insulated crimp-type terminal to the end of the cables.(1) Power cable, current output cable, digital I/O cablesThe necessary cables should be ordered from the person responsible for the installation. Strip the sheath ofeach conductor as shown in Figure 5.1 and attach a crimping terminal with insulated sleeve to it. The sizeof the crimping terminal is as follows:M4: Integral type LF600 and LF610M3.5: Integral type LF620, Separate type LF602, LF612 and LF622• Connect the power cable to terminal blocks L1 and L2.• Connect the current output cable to terminal blocks + and -.• Connect the digital I/O cable to terminal blocks D1, D01, D02 and COM, as required.Figure 5.1 Terminal Treatment of Power Cable, Current Output Cableand Digital I/O cable(2) Excitation cableStrip the sheath from the end of each conductor as shown in Figure 5.2, attach an M3.5 crimping terminalwith insulated sleeve, and connect it to the terminal blocks X and Y. Connect the red conductor to terminalblock E.Figure 5.2 Terminal Treatment of Excitation CableCrimping terminal25~45mmM3.5 crimping terminalX BlackE RedY White25~45mmM4: LF600F and LF610FM3.5: LF602F, LF612F, LF620F and LF622FSeparate