Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital12.05 CS SD 24Error messages of the imaging system 47Page 47 of 68Medical Solutions-519 RETRYCANCEL#1:Warning#2:External storage error:#3: The external storage memory capacity is insufficient. Retryafter making space on the disk.[Cause]There is not enough space on the external storage medium.[Operation after the reply key is pressed]When RETRY is selected, the image will be stored on the exter-nal storage medium.When CANCEL is selected, the storing ofthe image will be cancelled after a confirmation message is dis-played.-520 RETRYCANCEL#1:Warning#2:Internal storage error:#3:Exposed image cannot be stored, because the internal stor-age memory capacity is insufficient.[Cause]Space on the internal disk is insufficient.[Operation after the reply key is pressed]When RETRY is selected, the image will be stored on the inter-nal disk.When CANCEL is selected, the storing of the image willbe cancelled after a confirmation message is displayed.-6000 OK #1:Warning#2:Calibration error#3:STAND(SENSOR-ID:XXX) Time has elapsed since the lastcalibration. Calibrate now.[Cause]This error occurs when calibration has not been performedwithin the set time.[Operation after the reply key is pressed]The calibration screen appears to prompt you to perform calibra-tion.Code ReplykeysMessage