36 Error messages of the imaging systemMOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG12.05 CS SD 24Page 36 of 68 Medical Solutions-128 Detector Unit: Signal Line Error:Problem has occurred betweendetector unit and EO box. Call aservice person.• Touch OK and shut down the sys-tem.• If this error occurs, contact a Sie-mens representative or distributor.-129 Capture Board: Data TransmitError: Failed to transmit the data.Call a service person.• Touch OK to close the window.As image quality cannot be guaran-teed, take another exposure.-130 Capture Board Error: Detectedcapture board error.• Touch OK and shut down the sys-tem.• If these errors occur, contact a Sie-mens representative or distributor.-131 Capture Board Control Error:Detected capture board controlerror. Call a service person.-132 Capture Board Error: Detectedcapture board error. Call a ser-vice person.-133 Capture Board Access Error:Detected Communication Error.-134 Capture Board: Sequence Error:Failed to synchronize with thedetector unit. Call a service per-son.-135 Capture Board: Sequence Error:X-ray monitor in detector unitfailed to detect start of radiation.-136 Capture Board: Data TransmitError: Failed to transmit data.Call a service person.-137 Capture Board: Control LineError: Control line error hasoccurred in capture board. Call aservice person.-140 Acquisition Parameter Error:Detected illegal parameter error.-141 LUT Write Error: Quit the examand perform calibration.• Touch OK and calibrate the unit.-143 Calibration Table Error: Performcalibration.Error Code Error Message Remedy