Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital12.05 CS SD 24Error messages of the imaging system 17Page 17 of 68Medical Solutions-105 OK/ -97 #1:Fatal error#2:CXDCAP.INI Open error#3:Install CXDCAP.INI file.[Cause]The cxdcap.ini file has not been installed.[Operation after the reply key is pressed]The system enters a fatal condition. All the sensor units willtherefore be disconnected. A service person needs to be calledin. However, you can transfer the image and use the study list.-97Cxdcap.ini file not found.-522 OK #1:Fatal error#2:File Open Error:filename#3:File is not found.[Cause]For some reason, XXXX.BIN and XXXX.DCM files do not exist.Or, they cannot be read.[Operation after the reply key is pressed]Shutdown starts. Although this problem can be solved using theservice tool, users must not use the service tool on their own.Take necessary actions in accordance with the service person"sinstruction.Code ReplykeysMessage