EN-18 5000 Series Indicators3.3.6 Geographical Adjustment FactorRefer to Table 3-2 and set the GEO factor that corresponds to your location.00 to 31NOTE: Only an authorized manufacturer’s representative or certified verification personnel may make thesechanges. Changing the geographical setting alters the calibration values.Advance to the next menu.3.3.7 End Calibration3.3.5 Calibration TestCalibration test is used to compare a known calibration weight against the stored span calibration data.NOTE: Calibration Test is always available (even when LFT is set to ON).When tESt is displayed, press the Yes button to initiate Calibration Test.The display flashes 0. With no weight on the scale, press the Yes button to record the current zero point.The display shows --t-- while the zero point is recorded.The display flashes the span calibration weight using the value from the last calibration. The example showstest weight of 30 kg.Place the specified test weight on the scale and press the Yes button.The display shows --t-- while the data is processed.The display flashes the actual difference between the calibration data and the test weight.The example shows a 0.010 kg difference. The result of the Calibration Test is printed.After 5 seconds, Calibration Test ends, the scale returns to the active weighing mode and displays the currentweight. •••www. GlobalTestSupply .comFind Quality Products Online at: sales@GlobalTestSupply.com