EN-44 CKW-55 CHECKWEIGHER- - - - - SPAN CAL - - - - -Cal: 40 kgOld: 39.99 kgDiff: 0.01 kgWt. Ref. _____________12/01/04 1:10:00 PMUSER ID 654321PROJ ID 111111SCALE ID 123456Name _______________- - - - - END - - - - -- - - - - CAL TEST - - - - -Test: 10 kgAct: 10.01 kgDiff: 0.01 kgWt. Ref. _____________12/01/04 1:30:00 PMUSER ID 654321PROJ ID 111111SCALE ID 123456Name _______________- - - - - END - - - - -- - - - - LIN CAL - - - - -Cal: 40 kgOld: 39.99 kgDiff: 0.01 kgWt. Ref. _____________12/01/04 1:10:00 PMUSER ID 654321PROJ ID 111111SCALE ID 123456Name _______________- - - - - END - - - - -4.1.5 GMP Calibration PrintoutIf any setting in the Print-Content-GMP menu (refer to Section 3.6.4 Print Menu) is turned on, GMP automatically prints thecalibration data after calibration or calibration test is completed. Other GMP data, if turned on, will also be printed as shown.Span Calibration PrintoutWhen performing a Span calibration with any GMP setting turned on, a printout isautomatically made after the procedure is completed.Linearity Calibration PrintoutWhen performing a Linearity calibration with any GMP setting turned on, a printout isautomatically made after the procedure is completed.Calibration Test PrintoutWhen performing a Calibration Test with any GMP setting turned on, a printout isautomatically made after the procedure is completed.