EN-36 CKW-55 CHECKWEIGHERNote: The normally closed relay condition is only active while the Indicator is powered on.When powered off or when power is removed, the relay condition returns to a normally opencondition. Restoring power to the Indicator will restore the closed condition of the relays ifselected.OUTPUT SEQUENCEThis menu item is used to define the relay outputs to be sequenced as the weight readingchanges from under / accept / over.Selections are:Normal = The previously enabled relay will be disabled as the next relay is enabled.Hold = The previously enabled relay will hold the same state as the next relay isenabled.CONTACTThis menu item is used to define the alteration of relay contacts. Selections are:S i mS i mS i mS i mS i m = Simultaneous. Relays open or close at the same time.BBM = Break-Before-Make contact style.MBB = Make-Before-Break contact style.Note: A 100 ms delay or over-lap is used for the BBM and MBB timing.ENDRepresents the end of the menu.3.6.7 GMP Data MenuThe GMP Data Menu enables the storage of identification and traceability parameters for audittrail purposes. The entered data are available for printing when turned on in the GMP Set menu3.6.4.Note: When there is no data entered for the User ID, Project ID, or the Scale ID, and the printingof these values has been enabled, the description text will still be printed with a long underscoreline following it.ResetThis menu item is used to reset the GMP Data menu to factory defaults.Selections are: NO, Yes.User IDThis menu item is used to enter the User identification via the numeric keypad (up to 6characters).Project IDThis menu item is used to enter the Project identification via the numeric keypad (up to 6characters).Scale IDThis menu item is used to enter the Scale identification via the numeric keypad (up to 6 characters).