KILOTECH INC.www.kilotech.com 1821. Enter Setup Mode:21.1 If need to set configuration parameters, set user parameters, calibrate the scale, set current date or time,test some hardware… It’s need to long press SETUP key to enter setup mode21.2 After Entering Setup Mode, the main menu item CONFIG will be shown first.21.3 In Setup mode, use to select wanted submenu and wantedmenu item, select wanted choice, set wanted number, confirm and save data, or exit this mode. Pleaserefer the “4.Operation Menu Structure”.22. UPS world ship:22.1. To interface the scale with UPS world ship set the following parameters:Enter USER menu, set following parameters, if RS232 is used, set these on COM1, if USB is used, setthese on COM2:USER-COM1(or COM2)-BAUD.RT = 4800USER-COM1(or COM2)-BYT.FMT = 7E1USER-COM1(or COM2)-LAYOUT = SCP-1222.2. Set scale port to NCI3835 in UPS WorldShip:23. Error codes and Troubleshooting25.1 Meaning of Symbols:25.1.1 0 ------ Zero is over the setting range25.1.2 0 ------ Zero point is below the setting range25.1.3 Ad ------ Signal to ADC is over max. range25.1.4 Ad ------ Signal to ADC is below min. range25.1.5 ------ Weight is over upper limitation, or display data is over limitation25.1.6 ------ Weight is below lower limitation25.1.7 EEP.E1 ------ CONFIG or CAL parameters are not correctly set25.1.8 EEP.E2 ------ USER parameter is not correctly set25.1.9 Lo.bAt ------ Battery voltage is lower than setting.25.1.10 CAP.- - - ---- Next displaying content is Capacity25.1.11 CAP.ER ----- Parameters about Capacity is not correct25.1.12 CAL.Px ----- Calibration on point(x)25.1.13 CAL.OFF ----- Calibration Seal Switch is on OFF position25.1.14 CAL.ON ----- Calibration Seal Switch is on ON position25.1.15 CAL.Er ----- Calibration error, maybe input data or loaded weight is incorrect, unstable, un-linear25.1.16 CAL.End ----- Calibration is end25.1.17 OFF ----- Power OFF the indicator25.1.18 STB.ER ------ Unstable time is lager than setting of USER-OTHER-NLD.RNG25.1.19 AC.xxxx ----- Accumulation times is xxxx