Calibrator Output Operations 2Keysight U1401B User’s and Service Guide 49Table 2-1 Default settings for the autoscan outputMode ±1.5000 V ±15.000 V ±25.000 mAStep Amplitude Time interval Amplitude Time interval Amplitude Time interval1 +1.5000 V 02 sec +15.000 V 02 sec +00.000 mA 02 sec2 +1.2000 V 02 sec +12.000 V 02 sec +04.000 mA 02 sec3 +0.9000 V 02 sec +09.000 V 02 sec +08.000 mA 02 sec4 +0.6000 V 02 sec +06.000 V 02 sec +12.000 mA 02 sec5 +0.3000 V 02 sec +03.000 V 02 sec +16.000 mA 02 sec6 +0.0000 V 02 sec +00.000 V 02 sec +20.000 mA 02 sec7 –0.3000 V 02 sec –03.000 V 02 sec +16.000 mA 02 sec8 –0.6000 V 02 sec –06.000 V 02 sec +12.000 mA 02 sec9 –0.9000 V 02 sec –09.000 V 02 sec +08.000 mA 02 sec10 –1.2000 V 02 sec –12.000 V 02 sec +04.000 mA 02 sec11 –1.5000 V 02 sec –15.000 V 02 sec +00.000 mA 02 sec12 +0.0000 V 00 sec +00.000 V 00 sec +04.000 mA 00 sec13 +0.0000 V 00 sec +00.000 V 00 sec +08.000 mA 00 sec14 +0.0000 V 00 sec +00.000 V 00 sec +12.000 mA 00 sec15 –1.5000 V 00 sec –15.000 V 00 sec +16.000 mA 00 sec16 +0.0000 V 00 sec +00.000 V 00 sec +20.000 mA 00 sec