Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 129Sorting Based on Measured Results (Comparator Function)Setting Up Sorting Judgment ConditionsStep 8. Enter the limit value using the entry keys. When you enter the value, thesoftkey labels change to unit labels (p, n, u, m, x1).• You can clear your selected upper limit value by pressing the CLEARsoftkey.• You can enter the upper limit value using the LOWx(-1) softkey.• You can clear all the limit values and sets the status to OFF for yourselected bin by pressing the CLEAR LINE softkey.Setting up AUX functionIf the secondary parameter limit range is ON, the sorting result varies when itexceeds the secondary parameter limit range, depending on the ON/OFF stateof the AUX function (Table 7-4).Setup procedureStep 1. Press [Meas Setup].Step 2. Press the LIMIT TABLE softkey.Step 3. Use the cursor keys to select the AUX field.Step 4. Use the following softkeys:Table 7-4 Sorting result when measured secondary parameter value exceeds limitrangePrimary parameter sortingresul t AUX BIN function Sorting resul tOne of BIN1 to BIN9 OFF OUT_OF_BINSON AUX_BINNot sorted to any BINs No relation OUT_OF_BINSSoftkey FunctionON Turns ON the auxiliary function.OFF Turns OFF the auxiliary function.