128 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance MeterSorting Based on Measured Results (Comparator Function)Setting Up Sorting Judgment Conditions7-Step 7. Enter the limit value using the entry keys. When you enter the value, thesoftkey labels change to unit labels (p, n, u, m, x1).• You can clear your selected lower limit value by pressing the CLEAR softkey.• You can enter the lower limit value using the HIGHx(-1) softkey.• You can clear all the limit values and sets the status to OFF for yourselected bin by pressing the CLEAR LINE softkey.Step 8. Use the cursor keys to select the BIN 1 HIGH field.Step 9. Enter the limit value using the entry keys. When you enter the value, thesoftkey labels change to unit labels (p, n, u, m, x1).• You can clear your selected upper limit value by pressing the CLEARsoftkey.• You can enter the upper limit value using the LOWx(-1) softkey.• You can clear all the limit values and sets the status to OFF for yourselected bin by pressing the CLEAR LINE softkey.Step 10. Repeat Step 6 to Step 9 using the cursor keys until you have entered the lowerand upper limit values for bin 9.Procedure to set up secondary parameter limit range (ON/OFF andlower and upper limit values)Step 1. Press [Meas Setup].Step 2. Press the LIMIT TABLE softkey.Step 3. Use the cursor keys to select the BIN 2nd (secondary parameter) field.Step 4. Use the following softkeys:Step 5. Use the cursor keys to select the 2nd LOW field.Step 6. Enter the limit value using the entry keys. When you enter the value, thesoftkey labels change to unit labels (p, n, u, m, x1).• You can clear your selected lower limit value by pressing the CLEAR softkey.• You can enter the lower limit value using the HIGHx(-1) softkey.• You can clear all the limit values and sets the status to OFF for yourselected bin by pressing the CLEAR LINE softkey.Step 7. Use the cursor keys to select the 2nd HIGH field.Softkey FunctionON Turns ON the bin function of secondary parameter.OFF Turns OFF the bin function of secondary parameter.