Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 127Sorting Based on Measured Results (Comparator Function)Setting Up Sorting Judgment ConditionsFigure 7-3 Example of gap between limit rangesProcedure to set up BIN1 to BIN9 (ON/OFF and lower and upper limitvalues)Step 1. Press [Meas Setup].Step 2. Press the LIMIT TABLE softkey.Step 3. Use the cursor keys to select the BIN 1 field.Step 4. Use the following softkeys:Step 5. Repeat Step 3 to Step 4 using the cursor keys until you have turned ON the bin9.Step 6. Use the cursor keys to select the BIN 1 LOW field.Table 7-3 Lower and upper limit values of limit ranges for Figure 7-3Absolute modeTolerance mode (reference value: 1 nF)Absolute PercentLowerlimitvalueUpperlimitvalueLowerlimitvalueUpperlimitvalueLowerlimitvalueUpperlimitvalueBIN1 1 nF 1.02 nF 0 nF 0.02 nF 0% 2%BIN2 0.97 nF 0.98 nF –0.03 nF –0.02 nF –3% –2%BIN3 0.99 nF 1.03 nF –0.01 nF 0.03 nF –1% 3%When the low limit value is greater than the upper limit, a warningmessage “Improper high/low limits” is d isplayed.Softkey FunctionON Turns ON the bin function.OFF Turns OFF the bin function.